8. Use the probe adjustment tool to adjust the probe to show a flat top and
bottom to the waveform. Remove the adjustment tool before looking at the
waveform. Repeat until waveform top and bottom are flat.
9. Repeat this procedure from step 2 - 8. for each probe connected to each
channel. You must also run this procedure if you move a probe from one
channel to another.
Probes and ground leads tip
When taking a
real
measurement, use
the shortest
possible ground
lead and signal
path to minimize
probe-induced
ringing and
distortion on the
measured signal.
Signal with a short length ground
lead
Signal with a long ground lead
Getting acquainted with the oscilloscope
TBS2000B Series Oscilloscopes User Manual 31