Video and General Performance Verication
Eye Gain (Option SD or HD
and E YE or PHY)
This test uses a
n 800 mV, 1 MHz square wave to check the Eye Gain. This
requires a 1 MHz square wave signal, provided by Tektronix, from the
TG2000 AVC1 module.
1. Press FACTORY.
2. Press CONFI
G, and then select Utilities > Calibration. Press SEL to start
calibration.
3. Connect the
AVG1 output to the SDI A input on the waveform monitor, and
select the 1 MHz square wave signal.
4. Terminate
the other SDI A connector with a 75 SDI terminator. (The
accuracy of this check depends directly on the accuracy of t his terminator.)
5. Press the
up or down arrow key if necessary to navigate the Eye Gain Adjust
SDI A calibration case. (The number and ordering of calibration cases varies
with installed options. If an Audio option is installed, press the up or down
arrow key repeatedly to move to the second page.)
6. Press SEL to begin this test.
7. Press the DISPLAY SELECT 1 button.
8. Posit
ion the display or use cursors to measure the waveform amplitude.
Check that the amplitude is between 760 mV and 840 mV.
9. Reco
rd this level in the test record for Eye Gain.
10. If the Eye Gain is outside the specication, perform the following adjustment
pro
cedure:
a. Press the DISPLAY SELECT 4 button to return to the Calibration menu.
b. Use the General knob or up and down arrow buttons to adjust the
displayed amplitude of the e ye waveform until it is 800 mV.
c. Press and hold STATUS, and then select Exit to exit calibration mode.
d. Repeat the Eye Gain test from step 1 and record the new values in the test
record for Eye Gain.
11. Record this level in the test record.
12. Option PHY only. Check that the Eye Amplitude readout in tile 2 is between
760 mV and 840 mV.
13. Repeat steps 3 through 12 for the SDI B input.
14. Press the DISPLAY SELECT 4 button to return to the Calibration menu.
15. Press and hold STATUS, and then select Exit. Press SEL to exit the
Calibration menu without saving.
Waveform Rasterizers Specications and Performance Verication 8 9

Table of Contents

Other manuals for Tektronix WVR6020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix WVR6020 and is the answer not in the manual?

Tektronix WVR6020 Specifications

General IconGeneral
BrandTektronix
ModelWVR6020
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals