Audio Performance Verification
Control or Setting Value
Horizontal Tr
igger Position
30%
Trigger Slope
Rising edge
Trigger Leve
l
0mV
10. Check that the amplitude of the waveform displayed on the oscilloscope is
between 0.9 V and 1.1 V and record the value in the test record.
11. Move the cable to each of the remaining AES B I/O connectors and repeat
previous step.
12. Disconnect the test setup.
AES Output Jitter
This test
measures jitter a t each AES output.
Typical Operation Check. This test checks for typical operation. Typical values
are not guaranteed characteristics and are listed in the test record as Pass / fail.
Waveform Rasterizers Specifications and Performance Verification 113