Example (output from HMP63):
?
HMP63 / 1.0.4
Serial number : H3640004
Batch number : T0001109
Sensor number : H0000322
Sensor model : Intercap
Order code : A12A0A2B0
Cal. date : 20120907
Cal. info : VAISALA/HEL
Time : 00:21:05
Serial mode : ANALOG
Baud P D S : 19200 N 8 1
Output interval: 1 S
Serial delay : 30
Analog delay : 10 S
Address : 0
Filter : 1.000
Ch1 output : 0 ... 1 V
Ch2 output : 0 ... 1 V
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
If the probe is in POLL mode, but a connection has not been opened using the OPEN
command, issue the ?? command.
??<cr>
More information
Set serial line settings (page 42)
6.6.2 View calibration information
Use the CDATE command to view the calibration date and CTEXT to view the calibration info
text. Date format for CDATE is YYYYMMDD.
CDATE<cr>
CTEXT<cr>
Examples:
cdate
Cal. date : 20150109
ctext
Cal. info : VAISALA/HEL
Chapter 6 – Serial line communication
39

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala HMP110REF and is the answer not in the manual?

Vaisala HMP110REF Specifications

General IconGeneral
BrandVaisala
ModelHMP110REF
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals