8. Troubleshooting
8.1 Analog output error notification
If the device is unable to measure due to an error, the analog output is set to an error level. The
default error level is 0 V, or another value predefined by the customer when ordering the
device.
You can change the analog output error level using the AERR command.
More information
Set analog output error indication level (page 50)
8.2 Solving typical problems
You can check the error message via the serial interface by using the ERRS command. If you
are unable to remove the errors, contact Vaisala. For contact information, see Technical
support (page 95).
Table 14 Troubleshooting table
Problem or message Likely causes and solutions
Measurement not working. Any of the following
errors active:
T meas error
RH meas error
Sensor(s) damaged or missing. Open the filter
and check.
Check the supply voltage of the probe.
Check the output mode of the probe using serial
line.
Humidity measurement appears to be wrong. Check for condensation on the probe and
sensor.
Use the L command to check the currently
applied calibration correction.
Calibrate and adjust the probe (HMP110).
Calibrate the probe and change the sensor if
necessary (HMP60).
HMP60 and HMP110 Series User Guide M211060EN-J
66