Pax-i3D Smart(PHT-30LFO) Installation Manual
23
2 Choosing an Installation Site
2.2
Specications for Electrical Installation
These specications are based on the
MEIGaN
(Medical electrical installation guidance notes).
Consult the companion manual for further information. :
Volume 3: Specification for Electrical
Installation.
2.3
Electrical Requirements
This equipment must be connected to a grounded outlet to fulll the safety
provisions specied in IEC 60364: the 2
nd
edition (2006).
It is mandatory that both PC and equipment use the same power line if
connected to an MPSO.
Whenever possible, use different power outlets for each device. If a
multiple portable socket outlet (MPSO) must be used, ensure that the PC
and equipment are connected to the same MPSO.
Use a dedicated power outlet for the power cord. Failure to do so may
result in unstable system operation caused by power uctuations.
It is strongly recommended that you install an AVR. An AVR (automatic
voltage regulator) maintains a constant voltage and allows for continuous
operation in the event of a power uctuation.
Power supply voltage AC100-240 V
● The input line voltage depends on the local electrical distribution system.
● Allowable input voltage uctuation requirement: ± 10 %.
Frequency
50/60 Hz
Phase
single
Power rating (maximum power consumption)
Max.2.2 kVA (during exposure)