Pax-i3D Smart(PHT-30LFO) Installation Manual
28
2 Choosing an Installation Site
2.7
Installing the Warning Lamp and Door Interlock Switch
Refer to Appendix
A
for a complete installation guide.
● This system can be equipped with a warning lamp and the door interlock switch which are
activated when the X-Ray is energized.
● The warning lamp and the door interlock switch are not included with the equipment.
● The warning lamp and the door interlock switch must be installed by a qualied technician.
2.8
Installing the Emergency Stop Switch
Refer to Appendix
B
for a complete installation guide.
● Install the emergency stop switch along the main power cable in the central distribution panel.
● Install this switch so that it is within easy reach of the operator but cannot be accidentally
pressed.
● The switch must be a fool-proof model.