Option to install
Radio CO2 SpO2
SureTemp
Plus Printer Procedure
Install the printer.
Install the top housing.
Install the bottom housing.
Install the NIBP module.
Install the SpO2 module.
Install the CO2 module.
Install the communications board and radio.
Install the radio antenna.
Re-install the power-supply cover.
Re-install the rear housing.
Insert the battery.
After installing an option, be sure to confirm the proper operation of the monitor and its
options as described in “Functional verification tests.”
Masimo parameter upgrades
Masimo parameter upgrades are available for device models 6400, 6500, 6700, and
6800. Available upgrades are as follows:
• Hemoglobin (SpHb)
• Acoustic respiration (RRa)
Host firmware requirements
This section lists the host firmware requirements for each parameter. If your device does
not meet the minimum requirements, you can upgrade the host firmware online through
the Welch Allyn Service Tool.
• SpHb requirement: host firmware version 1.70.03 or later
• RRa requirement: host firmware version 2.00.00 or later
Upgrading devices configured without Masimo SpO2
You can upgrade devices configured without Masimo SpO2 by adding Masimo SpO2 or
by replacing the Nellcor SpO2 module with a Masimo SpO2 module.
122
Options, upgrades, and licenses Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™