Recommended service intervals
To confirm that the device is functioning within the design specifications, perform
periodic service as indicated in the following table. Customers who have the Silver
edition of the Welch Allyn Service Tool can perform the basic functional verification and
calibration procedures referenced in the table by following the instructions in this
manual. If you have the Gold edition of the service tool, use the tool to perform a
complete functional verification and calibration of the device in lieu of performing the
basic tests.
Component Service interval Service procedure
NIBP module Annually Basic functional verification
SpO2 module Annually Basic functional verification
SpHb parameter Annually Basic functional verification
CO2 module Annually or 1200 hours (whichever
comes first)
Calibration (first time only)
Annually or 4000 hours (whichever
comes first)
Calibration
30,000 hours Replace the module
SureTemp Plus Annually Basic functional verification
Braun ThermoScan PRO 4000 Annually Basic functional verification
Battery 300 charge cycles Replace the battery
Use the service tool, Gold edition, to perform a complete functional verification and
calibration of the device whenever any of the following conditions exist:
Based on the basic functional verification, the device does not meet specifications
The device has been dropped or otherwise damaged
The device is malfunctioning
The case has been opened
An internal part has been replaced (battery excluded)
Note For instructions on using the Gold edition, see the service tool help files.
Related Links
• “Basic functional verification tests” on page 102
Maintenance
For device maintenance information, see “Maintenance and service” in the device’s
directions for use. Covered topics include the following:
Replacing the printer paper
Inspecting and cleaning the device and accessories
14
Overview Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Welch Allyn Connex 6000 Series and is the answer not in the manual?

Welch Allyn Connex 6000 Series Specifications

General IconGeneral
BrandWelch Allyn
ModelConnex 6000 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals