Page #160 background imageLoading...
Page #160 background image
Factory defaults
General alarm
Settings Default value
Display alarm limits Enabled
Alarm audio on Enabled
Volume Medium
Advanced
Allow user to disable alarms Enabled
Allow user to turn off general audio Enabled
Minimum alarm volume Low
Nurse call threshold Medium
Audio pause time 2 minutes
SpO2 alarm condition delay 10 seconds
SpHb alarm condition delay 10 seconds
etCO2 alarm condition delay 15 seconds
Respiration alarm condition delay 10 seconds
No breath detected alarm delay
Adult: 30 seconds
Pediatric: 20 seconds
Neonate: 15 seconds