EasyManuals Logo
Home>YOKOGAWA>Voice Recorder>DX1012

YOKOGAWA DX1012 User Manual

YOKOGAWA DX1012
75 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
30 IM 04L41B01-02E
Setting the Data Storage
Setup Example 5: Continuously Record Measured Data and Automatically Save
In this example, we will continuously record and save the measured data of channel 1 and 2.
For the procedure to set the channel, see “Setting the Input Range and Alarm” on page 26.
For the procedure to set groups, see “Setting the Display” on page 28.
Automatically save to
the CF card periodicall
y
Channel 1
Channel 2
Setup Item Description Number in the Figure
Source channels Channel 1 (TI-001) and channel 2 (FI-002) 1
Data to be recorded Continuously record the display data from 2
the start of the measurement.
Data storage method Automatically store every 24 hours. 3
Sampling interval 4 seconds. Set using the trend update interval. 4
Sampling interval = (trend update interval setting)/30 dots
Data file name Add “sample” to the file name. 5
Example: 000123_sampleYYMMDD_HHMMSS0.DAD
↑↑
Sequential number Date/Time of the first recorded data and 0 (zero)
Save Destination Directory
DATA-101 6
(1) Data to be Recorded
Press MENU (switch to the setting mode), hold down FUNC for 3 s (switch to the
basic setting mode)
From the basic setting menu, select: A/D, Memory
2
(2) Method of Storing to the CF Card (Auto Save ON/OFF)
From the basic setting menu, select: Environment > Security, Media save
3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DX1012 and is the answer not in the manual?

YOKOGAWA DX1012 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDX1012
CategoryVoice Recorder
LanguageEnglish

Related product manuals