<2. WIRING AND INSTALLATION>
2-6
IM 12A01A02-01E 8th Edition : Oct. 01, 2015-00
F0203.ai
Cable gland nut
O-ring
Gable gland
O-ring
Gable gland
Sleeve
(for grounding cable line
of the plastic housing)
FLXA21 Plastic Housing
Cable gland cap
Cable gland nut
Gaskets
Washer
Cable gland nut
Gaskets
Washer
FLXA21 Stainless Steel Housing
FLXA202 Housing
Gable gland
Cable gland cap
Rubber plug
(without wiring)
Cable gland cap
Close up plug
(rubber)
Figure 2.6 Cable glands
The unused cable glands should be sealed with the supplied close up plug.
CAUTION
When installing cable glands, hold cable glands and tighten cable gland nuts to a torque of 6
N•m. If cable glands, not gland nuts, are tightened, O-rings may be come out from the proper
positions.