EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 113
Editing or deleting QAL3 values
Editing an individual QAL3 value
1 Click the number of the QAL3 value (column "No.") to be edited in the
control chart display (see page 112) in the "QAL3 Monitoring" page.
2 The following data and values can be changed in the "Edit data" page:
• "Technician",
• "Comment",
• "sAMS" (S
AMS
value for zero point or reference point).
3 Exit the page by clicking "Save" or "Cancel".
Deleting an individual QAL3 value
1 Click the number of the QAL3 value (column "No.") to be deleted in the
control chart display (see page 112) on the "QAL3 Monitoring" page.
2 Click "Delete" on the "Edit data" page.
3 Acknowledge the subsequent query (Are you sure you want to delete
this data?) by clicking "OK".
The data is deleted, and the control chart is displayed again.