EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 67
Manual calibration: Calibration method
Menu path
Setup Calibration Data Manual Cal. Method
The calibration method can also be set in the configurator (see page 63).
Calibration method
Analyzer Calibration method Remark
Uras26 Zero-point/end-point calibration cannot be altered
Magnos206, Magnos28 Zero-point/end-point calibration
or
substitute gas calibration
or
single-point calibration
If a substitute gas component has been
installed in the Magnos206 (see page 73)
or Magnos28 (see page 75) analyzer, the
substitute gas calibration is set ex works
as the calibration method; otherwise, the
zero-point/end-point calibration is set.
The single-point calibration can be se-
lected as an alternative.
Caldos27 Zero-point/end-point calibration
or
single-point calibration
set ex works
with standard gas (see page 77)
Caldos25 Zero-point/end-point calibration
or
substitute gas calibration
set ex works
can be selected if a substitute gas com-
ponent has been set up in the Caldos25
(see page 79) analyzer