EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A 19
Requirements at the installation site, power supply
Installation site
The gas analyzer is only intended for installation indoors; it may not be in-
stalled outdoors.
The installation site must be stable enough to bear the weight of the gas
analyzer!
Short gas paths
Install the gas analyzer as close as possible to the sampling location.
Install the gas conditioning and calibration modules as close as possible to
the gas analyzer.
Adequate air circulation
Provide for adequate natural air circulation around the gas analyzer. Avoid
heat build-up.
Protection from adverse conditions
Protect the gas analyzer from
• Cold
• Exposure to heat from e.g. the sun, furnaces, boilers
• Temperature variations
• Strong air currents
• Accumulation and ingress of dust
• Corrosive atmosphere
• Vibration.
Climatic conditions
Atmospheric pressure Atmospheric conditions
Relative humidity Max. 75 %, slight condensation permissible
Ambient temperature
EL3060-CU +5 to +50 °C
EL3060-Caldos25 +5 to +50 °C / +45 °C
1)
EL3060-Caldos27 +5 to +50 °C / +45 °C
1)
EL3060-Magnos206 +5 to +50 °C / +45 °C
1)
EL3060-Magnos28 +5 to +50 °C / +45 °C
1)
EL3060-Uras26 +5 to +45 °C
for storage/transport −25 to +65 °C
1) +45 °C when used together with EL3060-Uras26
Remark: The explosion protection is not impaired when the gas analyzer is
operated at temperatures below +5 °C to −20 °C. However, compliance with
the measurement data is not guaranteed in this temperature range.