88 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
Caldos27: Changing the range limits
Menu path
Setup Measurement Ranges
The range limits can also be set in the configurator (see page 82).
NOTES
Measuring ranges must not be set within ignition limits.
The calibration of the measuring range must be checked after the range
limits have been changed.
Range limits
The start value of the measuring range and the end value of the measuring
range must be within the reference measuring range.
Setting range
The measuring range can be set by the user within the limits applicable for
the sample component. It is permanently set to rTC = 0 to 64,000 for the
factory-set component "Relative thermal conductivity".
The largest measuring range is 0 to 100 % vol. or 0 % vol. to saturation, de-
pending on the measuring task.
Physical measuring range
The gas analyzer has one physical measuring range for each sample compo-
nent. The limits of this measuring range are normally 0 % vol. and 100 % vol.