96 EL3060 CONTINUOUS GAS ANALYZERS | OI/EL3060-EN REV. A
Configuration of the Ethernet connection
Menu path
Setup Device Settings Ethernet
Function
The Ethernet connection enables communication
• with the test and calibration software TCT-light,
• with the configuration software ECT,
• for transmission of the QAL3 data, if the option QAL3 monitoring has
been integrated in the gas analyzer,
• for reading the measured values and for calibrating and controlling the
gas analyzer via the Modbus TCP/IP protocol.
Parameters
The parameters which have to be input depend on the DHCP setting:
DHCP on: Network name (max. 20 characters, no blanks or special characters),
DHCP off: IP address, IP address mask and IP gateway address.
The network name can only be altered in the configurator (see page 53). The
default network name consists of "EL3K" and the last six positions of the
MAC address (example: "EL3KFF579A").
If the parameter "DHCP" is set to "off", the Ethernet configuration is reset to
the standard configuration (default IP address); this will prevent the inad-
vertent assignment of an IP address from a DHCP pool.
Addresses
The IP address, IP address mask and IP gateway address must be obtained
from the system administrator.
NOTE
The address bits that can be varied in the address mask may not all be set to
0 or 1 (broadcast addresses).
MAC address
The 12-digit MAC address is unique and stored in each device during manu-
facture. It cannot be altered.