Commissioning
CI/FEX300/FEX500-EN FEX300, FEX500 EN - 55
6.5.2 Parameterizing via the "Easy Set-up" menu function
The device can be factory parameterized to customer specifications upon request.
If no customer information is available, the device is delivered with factory settings.
The setting of the most current parameters is summarized in the "Easy Set-up" menu. This
menu provides the quickest way to configure the device.
The Easy Set-up menu allows you to select the language, engineering unit for flowrate, flow
range, totalizer unit, pulse/frequency mode, pulse per unit, pulse width, damping, status of
current output during alarm (Iout at Alarm, Iout Low Alarm, Iout High Alarm).
For detailed descriptions of these menus and parameters, see the chapter on the "Parameter
overview".
The following section describes parameterization via the "Easy Set-up" menu function.
Boiler 9
Q 0.00 l/min
Σ+ 0000.00 m³
Σ- 0000.00 m³
4. Use
to switch to the configuration level:
Access Level
Read Only
Standard
Advanced
Back Select
5. Use
or to select "Standard".
6. Use
to confirm your selection.
Enter Password
**********
RSTUVWXYZ012345
Next OK
7. Use
to confirm your password. A password is not
available as factory default; you can continue without
entering a password.
Menu
Easy Setup
Exit Select
8. Use
or to select "Easy Setup".
9. Use
to confirm your selection.