EasyManuals Logo
Home>Agilent Technologies>Multimeter>34405A

Agilent Technologies 34405A User Manual

Agilent Technologies 34405A
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
106 34405A User’s and Service Guide
4 Performance Tests and Calibration
6 Enter the actual applied input (see “Entering Adjustment
Values on page 91).
7 Press to start the adjustment. The CAL
annunciator flashes to indicate the calibration is in
progress.
Successful completion of each adjustment value is
indicated by a short beep and the primary display
briefly showing PASS.
An adjustment failure is indicated by a long beep, the
primary display showing FAiL and a calibration error
number appearing in the secondary display. Check the
input value, range, function, and entered adjustment
value to correct the problem and repeat the adjustment
step.
8 Repeat steps 3 through 7 for each gain adjustment point
shown in the table.
9 Verify the Frequency Gain adjustments using the
“Frequency Gain Verification Test on page 81.
NOTE
Always complete tests in the same order as shown in the appropriate
table.
Ta ble 2 0 Frequency Gain Adjustment
Input Voltage Frequency Function
Voltage as 1Vrms
Cal Item
1Vrms 1kHz Frequency 1kHz
Disp
34405A users guide.book Page 106 Saturday, September 2, 2006 3:38 PM

Table of Contents

Other manuals for Agilent Technologies 34405A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34405A and is the answer not in the manual?

Agilent Technologies 34405A Specifications

General IconGeneral
BrandAgilent Technologies
Model34405A
CategoryMultimeter
LanguageEnglish

Related product manuals