98 34405A User’s and Service Guide
4 Performance Tests and Calibration
9 Verify the DC Voltage Gain adjustments using the “DC
Voltage Gain Verif ication Test on page 78.I
DC Current Gain Adjustment Procedure
Review the “Test Considerations on page 71 and “Gain
Adjustment Considerations” on page 95 sections before
beginning this procedure.
1 Press to enter DC Current Gain Calibration.
2 The primary display will show the uncalibrated value and
the secondary display will show the reference value of Cal
Item.
3 Configure each Cal Item shown in the adjustment table
below.
4 Use (Auto) or (Range) to select the Cal
Item.
Ta ble 1 5 DC Voltage Gain Adjustment
Input Function Cal Item
Dual Banana Plug with copper wire short between 2 terminals DC Voltage Short
100 mV 100 mV
+ 1 V + 1 V
- 1 V - 1 V
10 V 10 V
100 V 100 V
1000 V 1000 V
Caution: Set the calibrator output to 0V before disconnecting from the multimeter input terminals.
DCI
NOTE
If the zero adjustment procedure has been recently performed prior to the
DC Current gain calibration procedure, the Cal Item 'Open' can be
neglected.
34405A users guide.book Page 98 Saturday, September 2, 2006 3:38 PM

Table of Contents

Other manuals for Agilent Technologies 34405A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34405A and is the answer not in the manual?

Agilent Technologies 34405A Specifications

General IconGeneral
BrandAgilent Technologies
Model34405A
CategoryMultimeter
LanguageEnglish

Related product manuals