46 34405A User’s and Service Guide
2Features and Functions
Storing and Recalling Instrument States
You can save and recall complete instrument states including
all front panel settings, all math registers, all Utility Menu
settings, and all bus specific settings. There are four user
storage registers numbered 1 through 4. An additional state,
state 0, is managed by the instrument and stores the last
power- down state. The instrument automatically saves the
complete instrument configuration to State 0 whenever a
power- down event occurs.
For remote operation, refer to the MEMory Subsystem, the
*SAV, and *RCL commands in the Agilent 34405A Online
Programmer’s Reference online help.
Storing a State
Before storing an instrument state, select the measurement
function, range, math operations, and so on, that you want
saved as a state. To store the instrument state:
1 Press , the display Store and Recall annunciators
will begin flashing.
2 Press or until only the Store annunciator is
flashing.
3 Press again.
4 Press or until the state number (1- 4) you
want to use is shown flashing in the secondary display.
5 Press to store the state. The secondary display
briefly shows donE when the state is successfully saved.
NOTE
The store function in the utility menu must be enabled (On) before you can
store states. Refer to “Using the Utility Menu” on page 41 for details.
Store
Recall
Store
Recall
Store
Recall
Store
Recall
Store
Recall
Store
Recall
34405A users guide.book Page 46 Saturday, September 2, 2006 3:38 PM