70 34405A User’s and Service Guide
4 Performance Tests and Calibration
Recommended Test Equipment
The test equipment recommended for the performance
verification and adjustment procedures is listed below.If the
exact instrument is not available, substitute calibration
standards of equivalent accuracy.
A suggested alternate method would be to use the Agilent
3458A 8½ - Digit Digital Multimeter to measure less accurate
yet stable sources. The output value measured from the
source can be entered into the instrument as the target
calibration value.
Ta ble 3 Recommended Test Equipment
Application Recommended Equipment Recommended Accuracy Requirements
Zero Calibration Shorting Plug--Dual banana plug with copper
wire short between the two terminals
DC Voltage Fluke 5520A <1/5 instrument 1 year spec
DC Current Fluke 5520A <1/5 instrument 1 year spec
Resistance Fluke 5520A <1/5 instrument 1 year spec
AC Voltage Fluke 5520A <1/5 instrument 1 year spec
AC Current Fluke 5520A <1/5 instrument 1 year spec
Frequency Fluke 5520A <1/5 instrument 1 year spec
Capacitance Fluke 5520A <1/5 instrument 1 year spec
34405A users guide.book Page 70 Saturday, September 2, 2006 3:38 PM

Table of Contents

Other manuals for Agilent Technologies 34405A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34405A and is the answer not in the manual?

Agilent Technologies 34405A Specifications

General IconGeneral
BrandAgilent Technologies
Model34405A
CategoryMultimeter
LanguageEnglish

Related product manuals