Getting Started Tutorial 1
34405A User’s and Service Guide 23
Measuring Resistance
Seven Ranges: 100.000Ω, 1.00000 kΩ, 10.0000 kΩ, 100.000 kΩ, 1.00000 MΩ,
10.0000 MΩ, 100.000 MΩ
Measurement Method: two-wire ohms
Open-circuit voltage limited to < 5 V
Input protection 1000 V on all ranges (HI terminal)
Measuring AC (RMS) or DC Current up to 1.2A
Three AC Current or DC Current Ranges: 10.0000 mA, 100.000 mA, 1.00000 A
Shunt Resistance: 0.1Ω to 10 Ω for 10mA to 1A ranges
Input Protection: Front Panel 1.25A, 500V FH fuse for I terminal
Typical Display:
12A
rms
1000VDC
500Vpk
1.2A
rms
1.25A/500V FH
LO
CAT II (300V)
750VAC
HI
I
V
12A
Fused
Resistance
Test
Current
Range
ACI
DCI
Typical ACI Display:
12A
rms
1000VDC
500Vpk
1.2A
rms
1.25A/500V FH
LO
CAT II (300V)
750VAC
HI
I
V
12A
Fused
+
AC or DC Current Source
-
m A DC
m A DC
Range
m A AC
m A AC
Range
Typical DCI Display:
34405A users guide.book Page 23 Saturday, September 2, 2006 3:38 PM

Table of Contents

Other manuals for Agilent Technologies 34405A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34405A and is the answer not in the manual?

Agilent Technologies 34405A Specifications

General IconGeneral
BrandAgilent Technologies
Model34405A
CategoryMultimeter
LanguageEnglish

Related product manuals