Page #85 background image
Performance Tests and Calibration 4
34405A User’s and Service Guide 85
Optional AC Current Performance Verification Test
Configuration: AC Current (CONFigure:CURRent:AC)
1 Select the AC Current function.
2 Select each range in the order shown below. Provide the
indicated input voltage and frequency. Compare
measurement results to the appropriate test limits shown
in the table. (Be certain to allow for appropriate source
settling.)
[1] Select 5½- digit measurement resolution
Ta ble 1 2 Optional AC Current Performance Verification Test
Current Input Frequency Function
[1]
Range
Error from Nominal
1 year
10mA 20Hz AC Current 10mA ± 0.16mA
10mA 45Hz 10mA ± 0.16mA
10mA 1kHz 10mA ± 60µA
10mA 10kHz 10mA ± 0.22mA
1A 1kHz 1A ± 6mA
100mA 1kHz 1A ± 1.5mA
10mA 1kHz 1A ± 1.05mA
34405A users guide.book Page 85 Saturday, September 2, 2006 3:38 PM