24 34405A User’s and Service Guide
1 Getting Started Tutorial
Measuring AC (RMS) or DC Current up to 12A
• 10 Amp AC Current or DC Current Range
• Shunt Resistance: 0.01 Ω for 10A range
• Internal 15A, 600V fuse for 12A terminal
Measuring Frequency
• Five Ranges: 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 750.00 V. Range is
based on the voltage level of the signal, not frequency.
• Measurement Method: Reciprocal counting technique.
• Signal level: 10% of range to full scale input on all ranges
• Gate Time: 0.1 second or 1 period of the input signal, whichever is longer.
• Input Protection: 750V rms on all ranges (HI terminal)
DCI
ACI
Typical ACI Display:
12A
rms
1000VDC
500Vpk
1.2A
rms
1.25A/500V FH
LO
CAT II (300V)
750VAC
HI
I
V
12A
Fused
+
AC or DC Current Source
-
ADC
A DC
Range
AAC
A AC
Range
Typical DCI Display:
Freq
Freq
Typical Display:
12A
rms
1000 VDC
5 00Vpk
1.2A
rms
1.25A/500V FH
LO
CAT II (300V)
75 0VAC
HI
I
V
12A
Fused
Frequency
Source
Hz
Range
VAC