EasyManuals Logo
Home>Agilent Technologies>Multimeter>34405A

Agilent Technologies 34405A User Manual

Agilent Technologies 34405A
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
64 34405A User’s and Service Guide
3 Measurement Tutorial
Temperature Measurements
The multimeter measures temperature by measuring the temperature
sensitive resistance of 5 k
W
thermistors.
Thermistors consist of semiconductor materials and provide roughly 10
times the sensitivity of an RTD. Because they are semiconductors, their
temperature range is more limited, commonly to –80
o
C to 150
o
C.
Thermistors have highly non–linear temperature–resistance relationships;
therefore their conversion algorithms are more complex. Agilent
multimeters use the standard Hart–Steinhart Approximation to provide
accurate conversions.