Features and Functions 2
34980A User’s Guide 15
The Analog Bus relays are automatically opened and closed as required
during the scan to connect to the internal DMM for the measurement.
For example, all 2- wire measurements use the ABus1 (MEAS) relays
for 4- wire measurements, the ABus2 (SENS) relays are used in addition
to the ABus1 relays.
Each time you initiate a new scan, the instrument will clear the
previous set of readings from memory.
Front Panel Operation:
To configure the measurement parameters and add a channel to the
scan list, use the Channel (Configure) key.
To initiate a scan and store all readings in memory, press the
Scan (Measure) key. If you press the Scan (Measure) key with no
scan list defined, the instrument initiates a DMM- only measurement
(see “Stand-Alone DMM Mode below).
To stop a scan in progress, press and hold the Scan (Measure) key.
To view the readings in memory, use the
View key (the readings are not
erased when you read them).
Remote Interface Operation:
To define the list of channels to be included in the scan list, use the
ROUTe:SCAN command.
To configure the measurement parameters on the desired channels,
use the
CONFigure and SENSe commands.
To initiate a scan and store all readings in memory, use the
INITiate
or READ? command. Each time you initiate a new scan, the instrument
will clear the previous set of readings from memory.
To stop a scan in progress, use the
ABORt command.
To view the readings in memory, use the
FETCh? command
(the readings are not erased when you read them).
NOTE
You can use the READ? command in one of three forms depending on
which measurement mode you wish to use.
If you omit the optional <ch_list> parameter and a scan list is not
currently defined, the READ? command applies to the internal DMM.
If you omit the optional <ch_list> parameter and a scan list is
currently
defined, the READ? command performs a scan of the channels in the
scan list.
If you specify a <ch_list>, regardless of whether a scan list is currently
defined, the READ? command performs a “temporary” scan of the
specified channels (independent of the present scan list).

Table of Contents

Other manuals for Agilent Technologies 34980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34980A and is the answer not in the manual?

Agilent Technologies 34980A Specifications

General IconGeneral
BrandAgilent Technologies
Model34980A
CategorySwitch
LanguageEnglish

Related product manuals