60 34980A User’s Guide
2 Features and Functions
Remote Interface Operation: Use the following commands to select the
reading format.
FORMat:READing:ALARm ON
FORMat:READing:CHANnel ON
FORMat:READing:TIME ON
FORMat:READing:TIME:TYPE {ABSolute|RELative}
FORMat:READing:UNIT ON
Non-Sequential Scanning
By default, the instrument scans the list of channels in ascending order
from slot 1 through slot 8 (channels are reordered as needed). If your
application requires non- ordered scanning of the channels in the present
scan list, you can use the non- sequential scanning mode.
This feature is available from the remote interface only.
• The scanning mode applies to the entire mainframe and cannot be
selectively used on individual modules.
• When sequential scanning is enabled (default), the channels in the scan
list are placed in ascending order from slot 1 through slot 8. Duplicate
channels are not allowed. For example, (@2001,1003,1001,1003) will be
interpreted as (@1001,1003,2001).
• When sequential scanning is disabled (OFF), the channels remain in the
order presented in the scan list (see exception below). Multiple
occurrences of the same channel are allowed. For example,
(@2001,2001,2001) and (@3010,1003,1001,1005) are valid and the
channels will be scanned in the order presented.
• When you specify a range of channels in the scan list, the channels are
always sorted in ascending order, regardless of the scan order setting.
Therefore, (@1009:1001) will always be interpreted as 1001, 1002,
1003, etc.
• If you define a scan list with the sequential mode enabled and later
disable the mode, the scan list will not be reordered; however, the scan
list will be treated as a non- sequential list thereafter.
• If you have defined a scan list with the sequential mode disabled (OFF)
and later enable the mode, the channels will be reordered.

Table of Contents

Other manuals for Agilent Technologies 34980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34980A and is the answer not in the manual?

Agilent Technologies 34980A Specifications

General IconGeneral
BrandAgilent Technologies
Model34980A
CategorySwitch
LanguageEnglish

Related product manuals