62 34980A User’s Guide
2 Features and Functions
• Readings acquired during a Monitor are not stored in memory
(however, all readings from a scan in progress at the same time are
stored in memory).
• The
INITiate command stores readings in memory. Use the FETCh?
command to retrieve stored readings from memory (the readings are
not erased when you read them).
Remote Interface Operation: The following command retrieves stored
readings from memory (the readings are not erased).
FETCh?
Use the following commands to query the statistics on the readings stored
in memory for a specific channel or from the internal DMM. These
commands do not remove the data from memory.
The following command retrieves the last reading taken on channel 1 on
the module in slot 3 during a scan.
DATA:LAST? (@3001)
The following command clears the contents of statistics memory for the
selected channel.
CALC:AVER:CLEAR (@3001)
Use the following command to determine the total number of readings
stored in memory (all channels) from the most recent scan.
DATA:POINTS?
The following command reads and clears the specified number of readings
from memory. This allows you to continue a scan without losing data
stored in memory (if memory becomes full, new readings will overwrite
the first readings stored). The specified number of readings are cleared
from memory, starting with the oldest reading.
DATA:REMOVE? 12
CALC:AVER:MIN? (@3005)
CALC:AVER:MIN:TIME? (@3005)
CALC:AVER:MAX? (@3005)
CALC:AVER:MAX:TIME? (@3005)
CALC:AVER:AVER? (@3005)
CALC:AVER:COUNT? (@3005)
CALC:AVER:PTPEAK? (@3005)
Minimum reading on channel
Time minimum was logged
Maximum reading on channel
Time maximum was logged
Average of all readings on channel
Number of readings taken on channel
Peak- to- peak (maximum–minimum)

Table of Contents

Other manuals for Agilent Technologies 34980A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34980A and is the answer not in the manual?

Agilent Technologies 34980A Specifications

General IconGeneral
BrandAgilent Technologies
Model34980A
CategorySwitch
LanguageEnglish

Related product manuals