Getting Started 1
U3402A User’s and Service Guide 17
Making Measurements
The following pages show you how to make measurement connections and how
to select measurement functions from the front panel for each of the
measurement functions.
NOTE
After measuring a high voltage measurement of up to 1000 VDC, you are recommended
to wait for approximate two minutes before measuring a low-level measurement with
1 to 10 µV resolutions.
After measuring a high current measurement using the A input terminal, you are
recommended to wait for approximate ten minutes before measuring a low-level DC
measurements of volts, amperes, or ohms; to achieve accurate measurement. This is
due to the thermal voltages generated during the high current measurements that may
cause errors when measuring the low-level measurements.