Getting Started 1
U3402A User’s and Service Guide 25
Performing diode/continuity test
Testi ng di od es
The diode test measures the forward voltage of a semiconductor junction of
approximately 0.5 mA. The beeper will emit a single beep tone when the input
voltage is below +0.7 V (approximately 1.4 kW) and emits a continuous beep
tone when the input voltage is below 50 mV (approximately 100 W).
Measurements are displayed as shown below:
Measurement method: 0.83 mA ± 0.2% constant current source, open-circuit voltage
limited to < 5 V
Response time: 70 samples per seconds with audible tone
Gate time: 0.1 s or 1 period of the input signal, whichever is longer
Input protection: 500 V RMS on all ranges
Reading rate Measurement display
Slow 1.2 V range
Medium
2.5 V range
Fast
4.0 V range
NOTE
The measurement value will display OL (overload) when the voltage measured is
> 1.2 V at slow reading rate
> 2.5 V at medium reading rate
> 4.0 V at fast reading rate
CAUTION
Disconnect circuit power and discharge all high-voltage capacitors before testing
diodes to avoid damaging the multimeter.

Table of Contents

Other manuals for Agilent Technologies U3402A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3402A and is the answer not in the manual?

Agilent Technologies U3402A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3402A
CategoryMultimeter
LanguageEnglish

Related product manuals