EasyManuals Logo
Home>Agilent Technologies>Multimeter>U3402A

Agilent Technologies U3402A User Manual

Agilent Technologies U3402A
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #82 background imageLoading...
Page #82 background image
4Performance Test
68 U3402A User’s and Service Guide
Resistance Verification Test
Configuration: 2- wire W
1 Select the resistance function.
2 Select each range in the order shown in Table 4- 4. Provide the resistance
value indicated. Compare measurement results to the appropriate test
limits shown in Table 4- 4. (Be certain to allow for appropriate source
settling.)
Table 4 - 4 2-wire W verification test
Function Reading rate Input Range Error from nominal
one year
2-wire W Slow 0.000 W 120 W
±8 mW
[1]
0.00000 W 1.2 kW
±50 mW
[1]
0.0000 W 12 kW
±0.5 W
[1]
0.000 W 120 kW ±5 W
0.00000 W 1.2 MW ±50 W
0.0000 W 12 MW ±0.5 kW
0.000 W 120 MW ±8 kW
Medium 0.00 W 400 W
±50 mW
[1]
0.0000 W 4 kW
±0.3 W
[1]
0.000 W 40 kW
±3 W
[1]
0.00 W 400 kW ±30 W
0.0000 W 4 MW ±0.3 kW
0.000 W 40 MW ±3 kW
0.00 W 300 MW ±50 kW
Slow 100.000 W 120 W
±108 mW
[1]
1.00000 kW 1.2 kW
±0.85 W
[1]
10.0000 kW 12 kW
±6.5 W
[1]
100.000 kW 120 kW ±65 W
1.00000 MW 1.2 MW ±0.65 kW
10.0000 MW 12 MW ±30.5 kW
100.000 MW 120 MW ±3.008 MW

Table of Contents

Other manuals for Agilent Technologies U3402A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3402A and is the answer not in the manual?

Agilent Technologies U3402A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3402A
CategoryMultimeter
LanguageEnglish

Related product manuals