1Getting Started
18 U3402A User’s and Service Guide
Performing voltage measurements
AC voltage:
Five ranges:
Slow reading rate: 120.000 mV, 1.20000 V, 12.0000 V, 120.000 V, 750.00 V.
Medium reading rate: 400.00 mV, 4.0000 V, 40.000 V,400.00 V, 750.0 V.
Fast reading rate: 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 750 V.
Measurement method: AC coupled true RMS—measures the AC component with up to 400 VDC
bias on any range
Crest factor: Maximum 3:0 at full scale
Input impedance: 1 MW ± 2% in parallel with < 120 pF on all ranges
Input protection: 750 V RMS on all ranges
1 Press .
2 Connect the red and black test leads to the respective input terminals as
shown in Figure 1- 10.
3 Probe the test points and read the display. In auto- ranging mode, the
multimeter automatically selects the appropriate range and the
measurement is displayed.
Figure 1-10 ACV terminal connection and display
CAUTION
Ensure that the terminal connections are connected correctly before making any
measurement. To avoid damaging the multimeter, do not exceed the rated input limit.
ACV voltage
source
+

Table of Contents

Other manuals for Agilent Technologies U3402A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3402A and is the answer not in the manual?

Agilent Technologies U3402A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3402A
CategoryMultimeter
LanguageEnglish

Related product manuals