1Getting Started
20 U3402A User’s and Service Guide
Performing current measurements
Measuring AC (RMS) or DC Current in mA
Three AC current or DC current ranges:
Slow reading rate: 12.0000 mA, 120.000 mA, 1.20000 A
Medium reading rate: 40.000 mA, 120.00 mA, 1.2000 A
Fast reading rate: 40.00 mA, 120.0 mA, 1.200 A
Shunt resistance: 0.1 W to 10 W for 12 mA to 1.2 A range
Input protection: Front panel 1.25 A, 500 V FH fuse for one terminal
1 Press or .
2 Power off the measured circuit.
3 Connect the red and black test leads to mA input terminal as shown in
Figure 1- 12.
4 Probe the test points in series with the circuit
5 Power on the measured circuit and read the display.
Figure 1-12 ACI RMS or DCI (mA) terminal connection and display
+
ACI or DCI
current source