EasyManuals Logo
Home>Agilent Technologies>Multimeter>U3402A

Agilent Technologies U3402A User Manual

Agilent Technologies U3402A
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
5 Disassembly and Repair
76 U3402A User’s and Service Guide
Operating Checklist
Before returning your multimeter to Agilent for service or repair check the
following items:
Is the multimeter inoperative?
q Verify the power line voltage setting.
q Verify the power line fuse is installed.
q Verify that the power cord is connected to the multimeter and to AC line
power.
q Verify the front panel power switch is depressed.
See page 79.
Is the multimeter’s current input inoperative?
q Verify the current input fuse.