6.3.3 High PEEP Alarm
Property Description
Alarm text
High PEEP
Priority
Medium
Alarm condition A High PEEP alarm will be given when the measured
PEEP is 30% above the set value for more than 15 seconds
Possible cause
Blocked leakage port.
Too short expiratory time.
Changes in airway resistance and or compliance.
Malfunction of the exhalation valve.
Blocked exhalation valve.
Reset criteria
PEEP has gone below the alarm limit (lower than 30%
above the set value).
Ventilator action The Vivo 45 LS will continue treatment according to the
current settings.
Setting range
On
Off
114 Alarms
User manual
Doc. 007301 K-1