5.7.2 Using the ventilator with a Nurse Call System
The ventilator can be connected to a nurse call system using the nurse call cable. When
connected, the ventilator alarms will also be forwarded to the nurse call system.
5.7.2.1 Connect the ventilator to a Nurse Call System
1 Connect the nurse call cable at the
back of the ventilator.
2 Test the connection by triggering an alarm on the ventilator and verify that the
nurse call system activates.
5.7.3 Using the ventilator with the FiO
2
Sensor
The FiO
2
sensor can be used to monitor and store FiO
2
measurements. The FiO
2
sen-
sor measures the fraction of inspired oxygen in the air channel of the ventilator. The