5.5.2 Insp. Pressure
Item Description
Definition The Insp. Pressure setting is used to define the airway pressure during
the inspiratory phase.
Minimum/maximum working IPAP is limited/achieved by a software
control of blower speed vs. measured pressure.
Setting min 4 cmH
2
O
Setting max 50 cmH
2
O
(limited to 30 cmH
2
O above PEEP in Pediatric mode)
Setting
resolution
0.5 below 10 cmH
2
O, 1.0 above 10 cmH
2
O
Default
value
15 cmH
2
O
5.5.3 PEEP
Item Description
Definition The PEEP setting is used to define the airway pressure at the end of
the expiratory phase.
Setting min 2 cmH
2
O
Off (for patient circuits with active exhalation valve)
Setting max 20 cmH
2
O
For pressure ventilation modes: The max setting is also limited by
Insp. Pressure -2 cmH
2
O and Min Pressure -2 cmH
2
O.
Setting
resolution
0.5 below 10 cmH
2
O, 1.0 above 10 cmH
2
O
Default
value
5 cmH
2
O (Adult mode)
2 cmH
2
O (Pediatric mode)
How to Use the Vivo 45 LS 79
Doc. 007301 K-1 User manual

Table of Contents

Other manuals for Breas Vivo 45 LS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 LS and is the answer not in the manual?

Breas Vivo 45 LS Specifications

General IconGeneral
BrandBreas
ModelVivo 45 LS
CategoryMedical Equipment
LanguageEnglish

Related product manuals