Item Description
Definition The Target Volume setting defines the tidal volume that the ventilator
will aim for while ventilating the patient in a pressure mode. To aim
for the preset volume, the ventilator will adapt the Inspiratory Pres-
sure between two adjustable pressure limits: Min Pressure and Max
Pressure.
When Target Volume is active, the mode field on the ventilator display
will indicate "(TgV)".
Setting min Off
50 ml (Pediatric mode)
300 ml (Adult mode)
Setting max
2000 ml (Adult mode),
400 ml (Pediatric mode)
Setting
resolution
10 below 500 ml,50 above 500 ml
Default
value
Off
5.5.18 Tidal Volume
ITEM DESCRIPTION
Definition The Tidal Volume setting defines the volume that will be delivered by
the Vivo 45 LS for each breath when using volume control modes.
In VCV-SIMV mode, this setting is applicable to the mandatory
breaths that are controlled by the ventilator.
Setting min 300 ml (Adult mode)
50 ml (Pediatric mode)
Setting max 2000 ml (Adult mode)
400 ml (Pediatric mode)
Setting
resolution
10 ml below 500 ml
50 ml above 500 ml
How to Use the Vivo 45 LS 87
Doc. 007301 K-1 User manual