5.5.14 Exp. Trigger (Expiratory Trigger)
Item Description
Definition The Expiratory Trigger setting defines the moment when the ventila-
tor will cycle from the inspiratory to the expiratory phase.
Setting min 1 (10% decrease of peak flow)
Setting max 9 (90% decrease of peak flow)
Setting
resolution
1 (Setting 1 is the most sensitive and 9 is the least sensitive)
5.5.15 Max Insp. Time (Maximum Inspiratory Time)
Item Description
Definition The Maximum Inspiratory Time setting defines a maximum length
for each inspiration. If the Maximum Inspiratory Time is set to Off,
the length of the inspiration and/or minimum inspiratory time is
dependent on the set Expiratory Trigger.
Modes PSV, PSV(TgV)
Setting min 0.3 s
Setting max
5 s (Adult mode)
2 s (Pediatric mode)
Off
Setting
resolution
0.1 s
Default
value
Off
How to Use the Vivo 45 LS 85