EasyManuals Logo
Home>Breas>Medical Equipment>Vivo 45 LS

Breas Vivo 45 LS User Manual

Breas Vivo 45 LS
194 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #157 background imageLoading...
Page #157 background image
7.1.1 Main Unit
1 Switch off the Vivo 45 LS and disconnect the power supply.
2 Remove the patient circuit.
3 Disconnect all electric cables.
4 Clean the outside of the Vivo 45 LS using a lint-free cloth with a mild soap solution,
and/or ethanol 70% for surface disinfection.
5 Reconnect the patient circuit. Make sure all parts are dry before the ventilator is put
into operation.
7.1.2 Air Pathway Disinfection
The table below lists the parts that might get contaminated by exhaled gases or bodily
fluids during normal use or single fault condition.
Condition Parts
With bacteria filter
Patient circuit
EtCO
2
airway adapter (if used)
Bacteria filter
Without bacteria filter
Patient circuit
EtCO
2
airway adapter (if used)
FiO
2
sensor (if used)
Patient air outlet/Pneumatic unit
Air bypass unit
Blower/Inlet silencer
Air inlet with filters
In case of contamination, the internal air pathways of the Vivo 45 LS may be disinfected
up to 10 times by a maximum 60 minute long validated ozone gas process.
Low resistance bacteria filter, if used, should be replaced every 24 hours.
7.1.3 Patient Circuit
The patient circuit should be cleaned and replaced in accordance with the manufac-
turer’s instructions and care provider’s instructions, where applicable. For safety
information, read 2.4 Usage of Patient Circuit, page 26.
Cleaning and Maintenance 157
Doc. 007301 K-1 User manual

Table of Contents

Other manuals for Breas Vivo 45 LS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 LS and is the answer not in the manual?

Breas Vivo 45 LS Specifications

General IconGeneral
BrandBreas
ModelVivo 45 LS
CategoryMedical Equipment
LanguageEnglish

Related product manuals