6.3.7 High MV
i
Alarm (High Inspired Minute Volume Alarm)
Property Description
Alarm text
High MVi
Priority
Medium
Alarm condition A High Inspired Minute Volume alarm will be given when
the monitored inspired minute volume exceeds the set limit
for 15 seconds.
Possible cause
Mismatch between Breath Rate, Inspired Tidal Volume set-
tings and the alarm setting.
Increased Breath Rate.
Leakage around the mask or within one of the components
of the circuit.
Reset criteria
When inspired minute volume is below set alarm limits
Setting range
Adult mode: 1.0 to 40 l/min
Pediatric mode: 1.0 to 20 l/min
Off
Setting resolution
0.5 l/min
Ventilator action The ventilator will continue treatment with the same settings.