Ambient pressure range:
700 to 1100 mbar, corresponding to
~4200 meters (13800 feet) above sea
level to ~700 meters(2300 feet) below
sea level, at normal atmospheric
pressure.
As seen in the graph above, the ventila-
tor is unable to deliver set max pressure
at a very low ambient pressure.
Ingress Protection:
IP22
Solid particle protection: Hazardous parts are protected from touch by fingers and by
objects greater than 12 mm.
Liquid ingress protection: The protection withstands dripping water less than 15 degrees
from vertical.
The ingress protection has been tested by water drips equivalent to 3mm rain/minute
for 10 minutes (2.5 minutes for each tilting direction).
8.2.6 Other
Patient Circuit Leakage
Recommended leakage: 20 to 50 l/min at 10 cmH
2
O (leakage circuit)
Minimum leakage: 12 l/min at 4 cmH
2
O (leakage circuit)
Oxygen Inlet
Oxygen inlet port: Maximum flow: 15 l/min (medical oxygen). Oxygen coupling is
type CPC PMCD181032.
Start-up Time
Start-up from unpowered state: about 20 seconds.
Sound Power Level
Sound level at 10 cmH
2
O in CPAP mode: Less than 30 dB(A). Measured at 1 m.
Alarm sound level : Adjustable 50–80 dB(A), Measured at 1m. Tolerance: ± 5 dB(A).
Technical Specifications 171
Doc. 007301 K-1 User manual

Table of Contents

Other manuals for Breas Vivo 45 LS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 LS and is the answer not in the manual?

Breas Vivo 45 LS Specifications

General IconGeneral
BrandBreas
ModelVivo 45 LS
CategoryMedical Equipment
LanguageEnglish

Related product manuals