No Item
1
Bacterial filter: Inter-surgical Flo-Guard Breathing Filter, Part No.
1690001
2
Patient Circuit: Inter-Surgical 1.6m smoothbore tubing
with 22F connector, Part. No. 5016000
3
Mouthpiece: Inter-surgical NIV Angled Mouthpiece with NOTCH
22M/15F, Part. No. 1938000
1 Inspect the parts for damages.
2 Connect the filter to the patient air outlet on the ventilator.
3 Connect the patient circuit to the filter.
4 Connect the mouthpiece to the patient circuit.
4.5 Inspecting the Vivo 45 LS before Use
Inspection of Device
• Check that there is no visible damage.
• Check that the surface is clean.
Inspection of Cables
• Check that all cables are recommended by Breas.
• Check that the cables are undamaged.
• Check that the cables are properly connected.
Inspection of Placement
• The Vivo 45 LS shall be placed on solid flat surface below the patient level (see 4.2
Placing the Vivo 45 LS, page 45).
• Make sure that nothing can block the air inlet at the side.
50 Preparing the Vivo 45 LS for Use
User manual
Doc. 007301 K-1