6.3.15 Apnea Alarm
Property Description
Alarm text
Apnea
Priority High
Alarm condition An Apnea alarm will be given when no patient-triggered
breath is detected for the set period of time.
The Apnea alarm is only available if the Inspiratory trigger is
activated.
Possible cause
Patient stopped breathing.
Patient decreases spontaneous breathing.
Circuit disconnection.
Inspiratory Trigger is set too high.
Reset criteria
Inspiratory effort detected by the Vivo 45 LS.
Ventilator action The Vivo 45 LS will continue treatment according to the
current settings.
Setting range
5 to 60 s. (Non MPV mode)
15 to 900 s. (MPV mode)
Off
Setting resolution 5 s below 15 s.
15 s above 15 s.
MPV mode:
15 s below 60 s.
60 s above 60 s.
6.3.16 Disconnection Alarm
CAUTION!
No single alarm can reliably detect all disconnections due to the number of possible
combinations of therapy settings, circuit configurations and patient interfaces. To
verify that patient disconnection can be detected, including if the patient interface
becomes accidentally detached from the patient, it is advised to test the functionality
of the Disconnection Alarm upfront with the complete set-up as used during treat-
ment, including items such as filters, circuit, connectors, interface (mask, cannula
etc.)
126 Alarms
User manual
Doc. 007301 K-1

Table of Contents

Other manuals for Breas Vivo 45 LS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Breas Vivo 45 LS and is the answer not in the manual?

Breas Vivo 45 LS Specifications

General IconGeneral
BrandBreas
ModelVivo 45 LS
CategoryMedical Equipment
LanguageEnglish

Related product manuals