FiO
2
measurements will be stored in the data memory which can be downloaded to a
PC and viewed in Breas software.
Usage
Time
Operating temperature
10 to 40°C (50 to 104°F)
Operating pressure
700 to 1250 mbar
Expected operating life <3 years (in ambient air) or 500,000 Vol.% h,
whichever comes first.
Shelf life < 6 months (recommended)
CAUTION!
Note that the operating conditions for the FiO
2
sensor is different from the ventila-
tor system conditions. If the sensor is used outside its operating conditions the FiO
2
measurements might deviate.
5.7.3.1 Installing the FiO
2
Sensor
1 Place the ventilator so the bottom is accessible.
2 Remove the hatch for FiO
2
sensor.
Use a torx TX10 screwdriver.
98 How to Use the Vivo 45 LS
User manual
Doc. 007301 K-1