Functional Checks
4-12 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Functional checks
Overview
This section describes the functional checks used with the
Invenia ABUS 2.0. Functional checks are used to verify that the
product works as specified after performing maintenance
procedures. Functional checks may also be used during
troubleshooting.
Performance Functional Checks
Power On/Boot Up Functional Check
Confirm that the system:
Powers on when you press the On/Off/Standby control.
That the transducer moves across the Scan Head
Assembly.
That the software runs through its initialization.
And that each user type (User or Admin) can successfully
login to the system.
Power Shut Down Functional Check
Confirm that the system exits the application and powers down.
Transducer Functional Check
Ensure that the transducer:
Upon Power Up moves across the Scan Head Assembly.
That the lights on the Scan Head Assembly light up.
That the Transducer images as a scan is being performed.
That the Transducer generates a 3D Volume.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Invenia ABUS 2.0 and is the answer not in the manual?

GE Invenia ABUS 2.0 Specifications

General IconGeneral
BrandGE
ModelInvenia ABUS 2.0
CategoryMedical Equipment
LanguageEnglish

Related product manuals