Replacement Procedures
8-2 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Warnings and important information
Warnings
WARNING
Energy Control and Power Lockout for Invenia ABUS 2.0.
When servicing parts of the Invenia ABUS 2.0 where there is
exposure to voltage greater than 30 volts:
1. Follow LOCKOUT/TAGOUT procedures.
2. Turn off the Power Mains Switch.
3. Unplug the Invenia ABUS 2.0.
4. Maintain control of the Invenia ABUS 2.0 power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
6. Remove/disconnect the battery, if present.
Invenia ABUS 2.0 components may be energized.
WARNING
Because of the limited access to cabinets and equipment in the
field, placing people in awkward positions, GE has limited the
lifting weight for one person in the field to 35 LBS (16 KG).
Anything over 35 LBS (16 KG) requires 2 people.
WARNING
At least two people are needed when replacing casters
(wheels) or adjusting brakes.
WARNING
Avoid contacting used parts before disinfecting.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Invenia ABUS 2.0 and is the answer not in the manual?

GE Invenia ABUS 2.0 Specifications

General IconGeneral
BrandGE
ModelInvenia ABUS 2.0
CategoryMedical Equipment
LanguageEnglish

Related product manuals