System Setup
3-42 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Connections on the Rear Panel
NOTE: Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC
standards (e.g. IEC60950 for data processing equipment and
IEC60601-1 for medical equipment). Furthermore, all complete
configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everybody who connects additional
equipment to the signal input part or signal output part of Invenia
ABUS 2.0, configures a medical system, and is therefore
responsible that the Invenia ABUS 2.0 device complies with the
requirements of the valid version of IEC60601-1-1. If in doubt,
consult the technical service department or your local
representative for GE.
Connection from Invenia ABUS 2.0 to a DICOM destination on a network
You will need one network cable to connect the Invenia ABUS
2.0 to a wall outlet on the hospital’s network.
1. Connect one end of the cable to the Ethernet connectors on
Invenia ABUS 2.0.
2. Connect the other end of the cable to the wall outlet.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Invenia ABUS 2.0 and is the answer not in the manual?

GE Invenia ABUS 2.0 Specifications

General IconGeneral
BrandGE
ModelInvenia ABUS 2.0
CategoryMedical Equipment
LanguageEnglish

Related product manuals