System Setup
3-40 Invenia ABUS 2.0 – System Setup and Basic Service Manual
4700-0043-00 Rev. 4
Setup Warnings and Cautions
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE
OF OPENING THE AC GROUND LINE (I.E., METER’S
GROUND SWITCH IS OPEN), DON’T TOUCH THE
ULTRASOUND SYSTEM!
To prevent electrical shock, connect the unit to a properly
grounded power outlet. Do not use a three to two prong
adapter. This defeats safety grounding.
Do not wear the ESD wrist strap when you work on live circuits
and more than 30 V peak is present.